Keywords: LUNA-III™ Automated Cell Counter, NIST Traceable Particle Size Standards, Cell size measurement, Size gating
Cell size is a crucial parameter in biomedical research and applications like biomanufacturing, where it indicates cell health and quality. In workflows such as biomanufacturing, cell size is essential for assessing batch culture or source material in cell therapy. This makes reliable cell size determination in automated cell counting valuable. The LUNA-III™ offers standard cell size gating features common to all LUNA™ Family devices, known for their precise cell size measurements. It also provides methods to analyze and categorize cell populations based on size during and after counting. We demonstrate the accuracy and reliability of the LUNA-FX7™ in measuring cell size and its precise size gating ability using six different NIST Traceable Particle Size Standards and the 3T3 cell line.
Cell Lines and Reagents
3T3 cell lines were used for the experiments. 3T3 cells were cultured in DMEM with 10 % FBS and 1 %. For viability assessments, 0.4 % trypan blue (Cat No. T13101) was utilized. Additionally, six different NIST Traceable Particle Size Standards (9.98, 20.85, 30.04, 43.33, 50.00, and 57.43 µm) were employed to verify size measurement accuracy.
Accuracy with NIST Traceable Particle Size Standards
NIST Traceable Particle Size Standards provide an official, objective comparison to known measurements maintained by the National Institute of Standards and Technology. These standards were used to verify the accuracy of the LUNA-III™ for size measurement. The LUNA-III™ demonstrated an accuracy of 95.2 % or higher in determining particle size, confirming its precision for size measurement (Figure 1).
Cell Lines Size Measurement
To evaluate the accuracy of size measurements in the LUNA-III™, we prepared a mixture of six NIST Traceable Particle Size Standards. We adjusted the counting protocol to include cell sizes from 1 µm to 90 µm. After counting, a histogram of cell concentration versus cell size was generated, accurately categorizing the NIST Standards. Different size beads were separately gated using flexible size gating based on the histogram, which can be adjusted immediately after counting. This size gating allowed for precise tagging of target sizes, achieving over 98 % accuracy in size measurement.
Cell Lines Size Measurement
To evaluate the LUNA-III™’s ability to filter 3T3 cells (known size: about 18 µm) based on size, we applied specific size gating criteria across various ranges: 7 µm – 24 µm, 1 µm – 11 µm, 12 µm – 24 µm, and 16 µm – 20 µm (Figure 3). The images show the accurate detection and sizing of 3T3 cells within each gating range, with corresponding histograms illustrating the distribution of cell sizes. The LUNA-III™ effectively filters and categorizes 3T3 cells by size, ensuring reliable and precise measurements across different cell size ranges.
The LUNA-III™ Automated Cell Counter provides highly accurate and reliable cell size measurements. Using NIST Traceable Particle Size Standards, the LUNA-III™ demonstrated an accuracy of 95.2 % or higher. It effectively categorized mixed particle sizes and accurately gated 3T3 cells across various size ranges, achieving over 98 % accuracy.
The LUNA-III™ offers:
• Precise cell size measurement across various particle sizes and cell types.
• Reliable size gating for accurate categorization of cell populations.
• High reproducibility and consistency in measurements.
These capabilities make the LUNA-III™ a valuable tool for assessing cell size, health, and quality.